Leave Your Message
AIM-M10 Insulation Monitoring Device
Medical IT Isolated Power System

AIM-M10 Insulation Monitoring Device

Application scenario: Economic solution for isolating power supply systems.  

Installation method: DIN-rail. (Dimension:54X85X64(mm))

Main functions:

  • Insulation monitoring for medical IT systems, range:10~999kΩ
  • Adjustable response value for insulation monitoring
  • Load and temperature monitoring for IT system transformers
  • Adjustable load current and temperature response value
  • Self monitoring with automatic alarm
  • PE connection monitoring
  • DC 24V power supply for AID
  • One relay 
  • 1 RS485, Modbus-RTU protocol
  • Up to 20 fault messages records

Standard:  IEC 61557-8, IEC 60364-7-710, IEC 61326-2-4 

    General


    AIM-M300 series medical intelligent monitoring device adopts advanced micro-controller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitization and networking in one. It is suitable for insulation monitoring of isolation power system in Class 2 medical locations 2 such as operating room and intensive care unit.

    Dimension


    7237b79452822b1af24358055da20c19

    Product Parameter


     Technical Data

    Wiring


    24V, G for the auxiliary power supply, and L1, L2 are connected to the IT system (which can be connected to the two output terminals of isolation transformer). I0, I1 for the current transformer signal input, and T0, T1 as the temperature sensor signal input. KE, FE are the functional grounding wires, which should be connected to the on-site equip potential terminals by two independent wires.

    wiring1

    wiring2

    Installation


    installation.png

    Diagram
    Diagram

    四件套

    Application应用(菲律宾站)
    Architecture

    architecture-1

    CONTACT US

    Your Name*

    Phone Number

    Country

    Message*